University of Cape Coast Institutional Repository

Investigations of Staff Doses during Computed Tomography-Fluoroscopy Procedures Using Monte Carlo Simulation

Show simple item record

dc.contributor.author Gyekye, Prince Kwabena
dc.date.accessioned 2022-08-11T10:08:58Z
dc.date.available 2022-08-11T10:08:58Z
dc.date.issued 2019-04
dc.identifier.issn 23105496
dc.identifier.uri http://hdl.handle.net/123456789/8472
dc.description xvi, 155p:, ill. en_US
dc.description.abstract Radiation dose to staff during Computed Tomography (CT) fluoroscopy procedures has been investigated using Monte Carlo N-Particle code 6 (MCNP6). Four (4) CT machines located at Korle-bu Teaching Hospital, Sweden Ghana Medical Centre, Karlsruhe Hospital and FOCOS Orthopaedic Hospital respectively were modelled using SimpleGeo software. The patient (supine) and staff (standing) were represented both by Female Adult meSH (FASH) and Male Adult meSH (MASH) computational voxel phantoms. Seven scenarios consisting of staff use of protective equipment, CT gantry and staff positioning in the CT room, CT room size variation, the use of lead drape and patient cover on patient, the use of protective face mask and apron by staff, scatter radiation distribution in the CT room, and the effective estimation of staff effective dose were investigated for occupational staff protection. A systematic approach for validating CT scanners with unknown bowtie-filter was proposed and tested. The use of protective equipment, lead drape and patient cover, protective face mask and apron, diagonal positioning of CT fluoroscopy gantry, and staff standing away from walls of room were observed to reduce effective dose to staff by a factor of 4.5, 4.4, 2.8, 1.1 and 1.2 respectively. Sex based single and double TLD effective dose estimation algorithm was proposed. The staff dose reduction factors observed for applying protection measures for this study varied by a factor of 1.6 - 6.9 compared with other studies. The use of conventional protection equipment at all times is recommended and complemented with alternative one (that is, standing position) to further reduce doses. en_US
dc.language.iso en en_US
dc.publisher University of Cape Coast en_US
dc.subject Computed Tomography en_US
dc.subject Effective dose en_US
dc.subject Fluoroscopy en_US
dc.subject Monte Carlo en_US
dc.subject Photon en_US
dc.subject Staff Protection en_US
dc.title Investigations of Staff Doses during Computed Tomography-Fluoroscopy Procedures Using Monte Carlo Simulation en_US
dc.type Thesis en_US