University of Cape Coast Institutional Repository

Assessment of some selected conventional diagnostic x-ray facilities at Cape-Coast in the Central Region of Ghana

Show simple item record

dc.contributor.author Owusu-Banahene, J
dc.contributor.author Owusu, I
dc.contributor.author Awuah, Baffour
dc.contributor.author Darko, E.O
dc.date.accessioned 2021-10-06T09:59:57Z
dc.date.available 2021-10-06T09:59:57Z
dc.date.issued 2015-09-29
dc.identifier.issn 23105496
dc.identifier.uri http://hdl.handle.net/123456789/6129
dc.description 5p, ill. en_US
dc.description.abstract The aim of the present paper is to assess some of the factors affecting on quality assurance of some conventional X-ray facilities such as reproducibility of tube voltage, dose output, X-ray tube consistency, accuracy of kVp, and half value layer. Examinations of these factors were studied using noninvasive kV meter of RMI 24A Multi-function meter. The quality assurance tests of X-ray diagnostic examination are measured and compared with the international tolerance. The half value layer of the radiation beam is an essential component of the assessment program. It is a measure of the beam hardness which relates to the type and thickness of shielding required in the facility and it also determines to a reasonable extent how much soft radiation is present in a beam. This soft radiation is absorbed in the surface tissue and results in unnecessary radiation being absorbed by the patient. In addition, it also contributes to unnecessary scatter within the patient which detracts from the desired image. The half value layer of a machine should not change for the lifetime of a machine unless someone removes the filter and as such it should be measured at least during the installation of a new machine en_US
dc.language.iso en en_US
dc.publisher University of Cape Coast en_US
dc.subject X-rays en_US
dc.subject Quality assurance en_US
dc.subject Half value layer en_US
dc.subject Radiation beam and examination en_US
dc.title Assessment of some selected conventional diagnostic x-ray facilities at Cape-Coast in the Central Region of Ghana en_US
dc.type Article en_US